Main Feature
- Automatic closed-loop test cycleaccording to selected Standards.
- Dual user interface display and PC
- Accuracy Class 1 (EN) and Class A (ASTM) starting from 10%of full scale
- Special calibration starting from few kNavailable on request;
- Compatible with the new, intuitive and smartDATA MANAGER Software
- Adopt the latest ES Energy Saving technologiesfor reduction of power consumption and silent operation
- Double stage hydraulic pump with rapid approach:high throughput of accurate tests(up to 40 per hour)
- Soft platen-to-specimen contact andsmooth load rate control from the very beginning of the ramp
- Optional control of a second and third framefor single chambers testers
- Optional internal graphic printerincluding Load-Time plot
- Remote verification of settings and performances for mulfunctions debugging
- Universal compression testers for testing cement, mortar, resins and other materials
- Storage of up to 10 test profiles for each channel for quick test start
General Description
These Multipurpose compression / flexure testers have been designed for various applications in which a limited maximum load is requested, together with a large testing space. They are suitable to determinate compressive strength on lightweight concrete, soil-cement specimen, high strength cement and refractory specimens.
The large testing space is ideal for performing, with the suitable accessories, splitting tests and flexural tests on concrete specimens.
They are available in the two different versions:
single compression station, 500 kN or 600 kN cap.:
for compression tests on low strength specimens. The Class 1 accuracy can be extended from 1% of testing machine full scale with special calibration procedures (see Upgrading Option and Technical Specifications).
double station model, 600/15 kN or 500/15 kN cap.:
extends the possible applications to flexural tests on cement and compression tests on low strength specimens. The Class 1 accuracy can be extended with special calibration procedures (see Upgrading Option and Technical Specifications).
Frames and compression platens
High rigidity solid one piece steel frame. Spherical seat allows free alignment at the initial contact with the specimen. For compression platens see Technical Specifications. Pedestal not included. See Accessories.
Pilot, Automatic power and control system
Dual stage pump, centrifugal for fast approach and multi-piston for loading, DC motor, ES Energy Saving technology to reduce the power consumption and ensure silent operation. For detailed information see Additional information.
Hardware
132,000 points effective resolution, three channels, touch screen graphic display 240x128 pixel, sampling rate 50/sec, large storing capacity on USB pen drive, test data downloadable to PC, Ethernet port, real time management DATAMANAGER software (optional, see Accessories)
Firmware
Simultaneous display of load-specific load, actual load rate and load/time graph; LAN connection to PC; advanced memory management; multi coefficient calibration.
Recording facility of up to 10 test profiles for each channel including type of test (e.g. compression, flexural, indirect tensile), specimen size and shape, load rate, standard in use and other general information. Each one of the recorded test profile can be recalled automatically to save time
Safety features
Includes: Max. pressure valve to avoid machine overloading; piston travel limit switch; emergency stop button; front door and rear transparent fragment guard on the compression station. Fragment guard for flexure station is available as optional (see Accessories).
Technical Specification
| Single Chamber | Single Chamber | Single Chamber | ||
| Models 50- | C92C02 | C92C12 | C93C02 | |
| Capacity [kN] | 600 | 500 | 600 | |
| Load measurement sensor | Pressure transducer | Load cell | Pressure transducer | |
| Class 1 range | 60 kN ÷ 600 kN | 50 kN ÷ 500 kN | 60 kN ÷ 600 kN | |
| With 50-C0050/CAL | 6 kN ÷ 600 kN | 5 kN ÷ 500 kN | 6 kN ÷ 600 kN | |
| With 50-C0050/1CAL | - | 0.5 kN ÷ 500 kN | - | |
| Platens, dia. [ mm] | 165 x 30 | 165 x 30 | 216 x 50 | |
| Hardness [HRC] | 55 | 55 | 55 | |
| Flatness [mm] | 0.02 | 0.02 | 0.03 | |
| Ram travel mm | 50 | 50 | 50 | |
| Max vertical daylight [mm]* | 345 | 240 | 315 | |
| Horizontal daylight [mm] | 265 | 265 | 265 | |
| Overall dimensions (lxdxh) [mm] | 760x400x1100 | 760x400x1100 | 760x400x1100 | |
| Weight approx. kg: | 270 | 280 | 300 |
| Double Chamber | Double Chamber | Double Chamber | ||
| Models 50- | C92C22 | C92C42 | C93C22 | |
| Capacity [kN] | 600/15 | 500/15 | 600/15 | |
| Load measurement sensor | P.transducer/Load cell | Load cells | P.transducer/Load cell | |
| Class 1 range | 60 kN ÷ 600 kN 1.5 kN ÷ 15 kN |
50 kN ÷ 500 kN 1.5 kN ÷ 15 kN |
60 kN ÷ 600 kN 1.5 kN ÷ 15 kN |
|
| With 50-C0050/CAL | 6 kN ÷ 600 kN | 5 kN ÷ 500 kN | 6 kN ÷ 600 kN | |
| With 50-C0050/CAL5 | 0.75 kN ÷ 15 kN | 0.75 kN ÷ 15 kN | 0.75 kN ÷ 15 kN | |
| Platens, dia. [ mm] | 165 x 30 / 165 x 30 |
165 x 30 / 165 x 30 |
216 x 50 / 165 x 30 |
|
| Hardness [HRC] | 55 | 55 | 55 | |
| Flatness [mm] | 0.02/0.02 | 0.02/0.02 | 0.03/0.02 | |
| Ram travel mm | 50/30 | 50/30 | 50/30 | |
| Max vertical daylight [mm]* | 345/205 | 240/205 | 315/205 | |
| Horizontal daylight [mm] | 265 | 265 | 265 | |
| Overall dimensions (lxdxh) [mm] | 1010x400x1100 | 1010x400x1100 | 1010x400x1100 | |
| Weight approx. kg: | 340 | 350 | 370 |






Reviews